Definition of the checklist functions that are developed in Eubrewnet
The users with the right permissions have access to some functions to directly check some tests that can be do to Brewers. The root of the access functions is the same for all http://eubrewnet.aemet.es/eubrewnet/data/checklist. From this point users can select different function. As Example Check SI.
The SI check function tests if the Brewer Operators perform Sittings periodically. The Regional Brewer Calibration Center Europe has stablished a standard of one test every week. The check is OK (returns True as result) if the number of Sittings per week is greater or equal to one. It is possible to modify the reference to change the number of sittings per week.
Description of the SI Check Function:
There is a Checklist Table with a summary of the available functions to perform tests and their links as example. It has the extended documentation for each function.
There are common inputs for the functions as the parameter brewerid wich can be used in almost every function are listed in the common inputs section. A short description of the outpus can be found in the common outputs section.
The functions can be run building a complete checklist with all tests in EUBREWNET CHECKLIST which can be obtained from the Menu - User - Checklist - Check. The list of all predefined date periods is in the Default Date Ranges section.
For security purposes user authentication has been added to this tools and their use is registered.
Almost all functions receive the following inputs, too:
All functions return the following outputs:
Section | Function | Default Date Range (days) |
---|---|---|
Instrument Operation | Setup and Level | 28 |
Instrument Operation | Sitting | 28 |
Instrument Operation | Files Read | 15 |
Instrument Historic | HP/HG | 365 |
Instrument Historic | Run Stop | 365 |
Instrument Historic | Deadtime | 365 |
Instrument Historic | SL R6 Temp | 365 |
Instrument Historic | SL R5 Temp | 365 |
Instrument Historic | F5 Intensity | 365 |
Instrument Weekly | HP/HG | 15 |
Instrument Weekly | Deadtime Reference | 15 |
Instrument Weekly | Deadtime STD | 15 |
Instrument Weekly | Run Stop | 15 |
Instrument Weekly | AP | 15 |
Instrument Weekly | SL Current | 15 |
Instrument Weekly | SL Voltage | 15 |
Instrument Weekly | HT Voltage | 15 |
Instrument Weekly | SL R6 | 15 |
Instrument Weekly | SL R5 | 15 |
Instrument Weekly | SL intensity | 15 |
Instrument Weekly | SL Outliers | 15 |
SL Temperature and Filter Analysis | Weekly SL Temperature Linearity | 15 |
SL Temperature and Filter Analysis | Weekly SL R6 Temperature Dependence | 15 |
SL Temperature and Filter Analysis | Weekly Temperature Range | 15 |
SL Temperature and Filter Analysis | Monthly SL Temperature Linearity | 45 |
SL Temperature and Filter Analysis | Monthly SL R6 Temperature Dependence | 45 |
SL Temperature and Filter Analysis | Monthly Temperature Range | 45 |
SL Temperature and Filter Analysis | Monthly Filter Attenuation | 45 |
SL Temperature and Filter Analysis | Monthly Filter Linearity | 45 |
SL Temperature and Filter Analysis | Monthly Filter Corrections | 45 |
Wavelength | Sun Scan Performed | 45 |
Wavelength | Calc Step | 45 |
Wavelength | OSC | 45 |
Wavelength | SC Range | 45 |
Wavelength | HS | 15 |
Wavelength | HL | 15 |