Brewer diagnostic tool

In this document we describe the Brewer diagnostic system. All the code is available in the rbcce subversion repository.

The main component is a parser. This parser extracts information from the B and D files and then stores usefull information in status files and to database as well.

Application Requirements

In order to have a good indicator of the brewer's status, the next error types must be detected and then they should be condensed in a sets of bits that can be decoded again. Every single error type have associated a mask that set or unset a single bit in this status bit sets. In the beginning of the proyect we need 4 bytes to represent 32 error types.

  • Number of type error limitations: The more error we detect the more bits we need. At some point we will be run out error types and we need another byte to archive the upgrade. In order to avoid this problem we added a unique error code integer to have more than 2^64 error types. In the future we can get rid of this sets of bits because this requirement is a limitation itself.
Mask A Mask B Mask C Mask D Code Level Description
00000001 00000000 00000000 00000000 11 Critical Azimuth tracker zeroing failure
00000010 00000000 00000000 00000000 12 Critical Brewer failed to respond
00000100 00000000 00000000 00000000 13 Critical Filterwheel #3 reference not found
00001000 00000000 00000000 00000000 14 Critical HG failed
00010000 00000000 00000000 00000000 15 Critical Lamp not detected
00100000 00000000 00000000 00000000 16 Critical Lamp not on
01000000 00000000 00000000 00000000 17 Critical Micrometer is jammed
10000000 00000000 00000000 00000000 18 - Error code not used
00000000 00000001 00000000 00000000 21 Critical Micrometer reset attempted
00000000 00000010 00000000 00000000 22 Critical No intensity detected
00000000 00000100 00000000 00000000 23 Critical Reset failed
00000000 00001000 00000000 00000000 24 Critical Micrometer alignment trouble
00000000 00010000 00000000 00000000 25 Critical Zenith zeroing failure
00000000 00100000 00000000 00000000 26 Warning Illegal command
00000000 01000000 00000000 00000000 27 - Error code not used
00000000 10000000 00000000 00000000 28 - Error code not used
00000000 00000000 00000001 00000000 31 Info UV scan started
00000000 00000000 00000010 00000000 32 Info UV scan finished
00000000 00000000 00000100 00000000 33 Warning UV scan aborted
00000000 00000000 00001000 00000000 34 Info End of day started
00000000 00000000 00010000 00000000 35 - Error code not used
00000000 00000000 00100000 00000000 36 - Error code not used
00000000 00000000 01000000 00000000 37 - Error code not used
00000000 00000000 10000000 00000000 38 Critical user break or unbreak
00000000 00000000 00000000 00000001 41 Critical no data
00000000 00000000 00000000 00000010 42 Critical UTC date failure
00000000 00000000 00000000 00000100 43 Critical Julian date failure
00000000 00000000 00000000 00001000 44 Critical time failure
00000000 00000000 00000000 00010000 45 Error Manual sighting requested
00000000 00000000 00000000 00100000 46 Info Total ozone submited
00000000 00000000 00000000 01000000 47 Warning Unexpected schedule running
00000000 00000000 00000000 10000000 48 - Error code not used

There are another set of test that should be done comparing a parsed value against a threshold. This threshold could be obtained from 3 different ways per each test.

  1. default value: in the database we will store a minimum value and a maximum value.
  2. brewer dependant value: in the database we will store a minumum and maximum per brewer.
  3. brewer dependant from config: in the database we will get a value from the brewer config and then we sustract and add 2 values [config - valueA, config + valueB].

If the tested parameter is in this range the Critical Level should be INFO, otherwise should be WARNING.

Code Thereshold Type Thereshold Default Description
51 default [-5,5] Azimuth discrepancy
52 config [config - 5, config + 5] Azimuth steps per revolution
53 default [-2,2] Zenith discrepancy
54 default [-15,15] time correction
55 TBD TBD micro is moved
56 15?? 15?? dark count
57 TBD TBD micrometer #1 diode found
58 TBD TBD micrometer #1 reference found
61 NA* NA* running schedule

TBD: to be defined.

NA: not applicable (there are no value to be tested)

Error Information Extracted

Additionally to the information extracted stated above, we need to extract more information because we need to check if the error is a false positive in the sources of error. In addition, we could need create complex database querys to make a more accurate diagnosis.

The information that we extract from the files will be described below.

Field Description
'brewerid' unique identificator from which brewer the error comes
'date' date when the error was produced
'gmt' time in GMT when the error was produced taken from the files
'level' integer that represents the importance of the error
'code' unique error code for a given error type
'filename' the file's name where the error was extracted
'line' the line inside the file where the error was extracted
'message' the original message extracted from the file
'count' ocurrences of the same error type in this file until now

Errors pending to implement

The following errors: UTC date failure, Julian date failure and Time failure, are difficult to extract. The only way to know the status of the remote clock is with the files' modification time. When we are copying those files to the local machine, we lose the modification time and its imposible to check if the remote clock is shift.

Proposed solution: We need to create a mecanism that allows the status tool to pool the remote clock in order to check if is synced.

Manual sighting request

This problem is hard to parse because in the B or D files appears to be a user command. This means that the user can type a error message in their own language. Sometimes we found the message in english but almost the majority of times we found this messages in a different language which is imposible to parse.

Flow Graph

  1. Every 15 minutes the eubrewnet system pools the B and D files from the brewers.
  2. The diagnostic tool parser receives it as input.
  3. The output is written to database and files.
  4. The Eubrewnet webpage shows the information extracted before.

Front End

The front end consist of two views, the main view that is the brewer overview and the detailed view.

Main View

When the users get into the frontpage, they can see a map with the status of the brewers. A code of colours states the status of the brewer. green means OK, yellow means WARNING, orange means ERROR, red means CRITICAL, grey means no data since the last 24 hours and finally black means no data since the last 48 hours.

example:

Detailed View

When the user clicks in a brewer, the user will be redirected to the details of the brewer. In the Brewer Status section we can see a list of original messages taken from D or B files explaining the status of the brewer.

The next example shows a brewer in a OK status.

The next example shows a brewer with several CRITICAL problems.

Detailed operation

The entry point of the diagnosis tool is monitor.py. The pseudo code of monitor.py is the following:

  1. create a list of path to posible files
  2. test if this file exist
  3. parse the file
  4. write the status on file
  5. write the status on database

The third item on the previous list is the parser itself and the file that contain the parser is parseSingleFile.py

parseSingleFile is a parent class of parseDFile.py and parseBFile.py. It contains many functionality to handle files and database access.

ParseBFile.py and ParseDFile.py contains an abstract function called parse_lines() that takes a list of lines of a given file and then creates a list of error. The pseudocode is the following:

  1. iterate over all lines of a file
    1. iterate over all error types
      1. if the line contains a specific string
        1. create new error
        2. push this new error in the errorList grouping by error code.